Open original FDA record ↗Recall record
- FDA identifier
- Z-0805-2013
- Company in FDA record
- Aesculap, Inc.
- FDA classification date
- Feb 12, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in the ventricle. There is a small percentage of cases where aggressively attempting to adjust the deflector has a potential to cause the catheter
Source facts
- Action
- Aesculap sent an Important Recall Notification letter dated June 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return the contact their local sales representative or customer service department at 1-800-282-9000 to return the affected product. Customers were also instructed to complete the Inventory Sheet attached to the letter with the quantity being returned. If customers could not locate the affected product they should provide an explanation as to why the inventory will not be returned. Customers with questions should call 610-984-9251 or 610-984-9291. For questions regarding this recall call 1-800-234-9179, ext 5067.
- Initiated
- 2012-06-04
- Root cause
- Process control
- Terminated
- 2014-01-08
- Distribution
- Nationwide Distribution including MI, PA and TX.
- Recall number
- Z-0805-2013
- Recall status
- Terminated
- CDRH record ID
- 113098
- FDA classified
- 2013-02-12
- Recall event number
- 63244