Z-0805-2014: ORCHESTRA/ORCHESTRA PLUS Programmer

Recall record Z-0805-2014 for Sorin Group Italia S.r.l.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0805-2014
Company in FDA record
Sorin Group Italia S.r.l.
FDA classification date
Jan 22, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sorin has voluntarily issued a notification to physicians related to the overestimation of the residual longevity displayed by its programmer (ORCHESTRA or ORCHESTRA PLUS) during a follow-up exam of patients implanted with a REPLY or ESPRIT pacemaker.

Source facts

Action
On Nov 20 and Nov 25, 2013, the Dear Doctor Letter was sent by certified mail to 1,310 physicians who are following 13,003 patients implanted with a REPLU or ESPRIT pacemakers in the US. Sorin provided the following recommendations to the physicians: You should consider checking the battery impedance of the last follow-up exam. In case the battery impedance is greater than or equal to 3.5k, a follow-up visit must be scheduled within a maximum of 6 months from the last follow-up visit. When pacemaker operation is checked by the simple application of a magnet, a magnet rate less than 95 min-1 should trigger a follow-up exam in the pacemaker centre. As a general rule, a maximum of 6 month follow-up interval when the battery impedance becomes greater than or equal to 3.5 k. This recommendation should also be followed subsequent to the installation of the new programmer software version.
Initiated
2013-11-20
Root cause
Software design
Terminated
2014-09-08
Distribution
Nationwide.
Recall number
Z-0805-2014
Recall status
Terminated
CDRH record ID
124518
FDA classified
2014-01-22
Recall event number
67013

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗