Open original FDA record ↗Recall record
- FDA identifier
- Z-0805-2015
- Company in FDA record
- American Medical Systems, Inc.
- FDA classification date
- Dec 16, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During routine periodic packaging testing, AMS identified that in simulated extreme shipping and handling situations the sterile barrier packaging of the Needle Passer components supplied with the AdVance" Male Sling System; Model # 720088-01; AdVance" XP Male Sling System; Model # 720163-01; Monarc" C Subfacial Hammock; Model # 72404195; Monarc" + Subfacial Hammock; Model #
Source facts
- Action
- Consignees were sent an AMS "Urgent Recall Notice" dated October 15, 2014. The letter was addressed to Physician, Health Care Professional, Nurse and Risk Manger. The letter described the problem and the product involved in the recall. Requested consignees to return all affected product and to complete and return the Acknowledgement Form. For questions, they can contact American Medical Systems Customer Service at 1-800-328-3881 (x6469). A second letter was available as optional communication to implanting physicians known by AMS to be regular users of the device. This physician letter was dated October 20, 2014.
- Initiated
- 2014-10-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-05-28
- Distribution
- Worldwide Distribution-US (nationwide) including DC and Puerto Rico and the countries of AUSTRIA, BELGIUM, CROATIA, CECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, LEBANON, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SLOVENIA, SERBIA, SOUTH AFRICA, SPAIN. SWEDEN, SWITZERLAND, TURKEY, UNITED KINGDOM, CHILE, COLUMBIA, COSTA RICA, HONG KONG, MALAYSIA, MARTINIQUE, NEW ZEALAND, PANAMA, PHILIPPINES, SINGAPORE, THAILAND, VENEZUELA, CANADA, AUSTRALIA.
- Recall number
- Z-0805-2015
- Recall status
- Terminated
- CDRH record ID
- 131551
- FDA classified
- 2014-12-16
- Recall event number
- 69796