Open original FDA record ↗Recall record
- FDA identifier
- Z-0805-2019
- Company in FDA record
- ICU Medical, Inc.
- FDA classification date
- Feb 22, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
There is a potential for one lot of ChemoLock Vial Spike to contain burr particulate
Source facts
- Action
- On 01/03/2019, Stericycle, Inc. sent an " URGENT: MEDICAL DEVICE RECALL" notification of the recall to ICU Medical direct consignees of record via traceable mailing. Consignees who further distributed the product were asked to notify their customers. The notification described the product, problem and actions to be taken. The consignees were instructed to do the following: 1)Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7114 (M-F, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Please visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, ICU Medical will credit you for any product returned. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product/ Table 1 sections of this notification and ask them to contact Stericycle at 1-888-871-7114 (M-F, 8am-Spm ET) to obtain a response form. For further inquiries, including product replacement options, please contact ICU Medical using the information provided below. Global Complaint Management at 1-844-654-7780 or email to: (ProductComplaintsPP@icumed.com) To report adverse events or product complaints or
- Initiated
- 2019-01-03
- Root cause
- Vendor change control
- Terminated
- 2024-05-29
- Distribution
- US Distribution to states of: OH, FL, IA, CA, PA, IL, VA; and internationally to: AUSTRIA, CHINA (TAIWAN), IRELAND, ITALY and SINGAPORE.
- Recall number
- Z-0805-2019
- Recall status
- Terminated
- CDRH record ID
- 170359
- FDA classified
- 2019-02-22
- Recall event number
- 81942