Open original FDA record ↗Recall record
- FDA identifier
- Z-0805-2025
- Company in FDA record
- Qualigen Inc
- FDA classification date
- Jan 3, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
Source facts
- Action
- The recalling firm issued letters dated 11/19/2024 via email to their customers explaining the reason for recall and providing the correct expiration dates for each of the affected lot numbers. The risk to health was provided and instructions given to be completed. The instructions were to cease use of the affected product, dispose of all remaining kits from the affected lot numbers, document the number of kits disposed, complete the acknowledgment form by FAX or email within 5 days.
- Initiated
- 2024-12-03
- Root cause
- Labeling Change Control
- Distribution
- Distribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.
- Recall number
- Z-0805-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211243
- FDA classified
- 2025-01-03
- Recall event number
- 95807