Skip to content

Public FDA record

Z-0805-2025: The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measure

Recall record Z-0805-2025 for Qualigen Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0805-2025
Company in FDA record
Qualigen Inc
FDA classification date
Jan 3, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.

Source facts

Action
The recalling firm issued letters dated 11/19/2024 via email to their customers explaining the reason for recall and providing the correct expiration dates for each of the affected lot numbers. The risk to health was provided and instructions given to be completed. The instructions were to cease use of the affected product, dispose of all remaining kits from the affected lot numbers, document the number of kits disposed, complete the acknowledgment form by FAX or email within 5 days.
Initiated
2024-12-03
Root cause
Labeling Change Control
Distribution
Distribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.
Recall number
Z-0805-2025
Recall status
Open, Classified
CDRH record ID
211243
FDA classified
2025-01-03
Recall event number
95807

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.