Skip to content

Public FDA record

Z-0806-05: Welch Allyn AED 20 Automated External Defibrillator; a portable, 12V internal battery powered defibrillat

Recall record Z-0806-05 for MRL, Inc., A Welch Allyn Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0806-05
Company in FDA record
MRL, Inc., A Welch Allyn Company
FDA classification date
May 12, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Welch Allyn AED 20 product may display a 'Defib Comm' error message on the device display during use which may result in a terminal failure of the device to analyze the patient's ECG and deliver the appropiate therapy.

Source facts

Action
Urgent Medical Device Recall letters dated 04/28/2005 were sent to the direct accounts informing them of the possibility in a terminal failure due to the radio frequency interference generated inside the AED 20 resulting from an intermittent electrical short. Any questions were directed to Welch Allyn-MRL Technical Support at 1-800-462-0777.
Initiated
2005-04-28
Root cause
Device Design
Terminated
2006-09-01
Distribution
Nationwide and internationally to Australia, Canada, China, France, Germany, Ireland, Israel, Malaysia, the Middle East, New Zealand, the Philippines, Poland, Singapore, South America, South Korea, Switzerland and the United Kingdom.
Recall number
Z-0806-05
Recall status
Terminated
CDRH record ID
38777
FDA classified
2005-05-12
Recall event number
31934

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗