- FDA identifier
- Z-0806-05
- Company in FDA record
- MRL, Inc., A Welch Allyn Company
- FDA classification date
- May 12, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Welch Allyn AED 20 product may display a 'Defib Comm' error message on the device display during use which may result in a terminal failure of the device to analyze the patient's ECG and deliver the appropiate therapy.
Source facts
- Action
- Urgent Medical Device Recall letters dated 04/28/2005 were sent to the direct accounts informing them of the possibility in a terminal failure due to the radio frequency interference generated inside the AED 20 resulting from an intermittent electrical short. Any questions were directed to Welch Allyn-MRL Technical Support at 1-800-462-0777.
- Initiated
- 2005-04-28
- Root cause
- Device Design
- Terminated
- 2006-09-01
- Distribution
- Nationwide and internationally to Australia, Canada, China, France, Germany, Ireland, Israel, Malaysia, the Middle East, New Zealand, the Philippines, Poland, Singapore, South America, South Korea, Switzerland and the United Kingdom.
- Recall number
- Z-0806-05
- Recall status
- Terminated
- CDRH record ID
- 38777
- FDA classified
- 2005-05-12
- Recall event number
- 31934