Z-0806-2012: Olympus OFP Flushing Pump irrigation tubing The Olympus OFP Flushing Pump is a peristaltic flushing pump

Recall record Z-0806-2012 for Olympus America Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0806-2012
Company in FDA record
Olympus America Inc.
FDA classification date
Jan 18, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Water may leak from the irrigation tubing used on the Olympus Flushing Pump

Source facts

Action
The firm,Olympus, sent an "URGENT-DEVICE CORRECTION " letter dated December 12, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were informed that any current irrigation tubing in their inventory can be used with the OFP pump; however, they must inspect the tubing installed on the OFP pump for signs of water leakage; if leakage is found, discontinue use of the tubing and replace with a new irrigation tube. In addition, complete and return the enclosed questionnaire via fax to (484) 896-7128. If you do not wish to use your current inventory of irrigation tubing and would like a replacement, call Olympus Customer Care Center at 1-800-848-9024. Olympus is discontinuing the current version of irrigation tubing and has redesigned the irrigation tubing to prevent water leakage. For any additional information, contact the V.P., Regulatory/Clinical Affairs & Quality Assurance at (484) 896-5688 or email at laura.storms-tyler@olympus.com.
Initiated
2011-12-12
Root cause
Nonconforming Material/Component
Terminated
2012-09-25
Distribution
Nationwide distribution: USA including states of: AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, AND WY.
Recall number
Z-0806-2012
Recall status
Terminated
CDRH record ID
105979
FDA classified
2012-01-18
Recall event number
60613

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗