- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Water may leak from the irrigation tubing used on the Olympus Flushing Pump
Source facts
- Action
- The firm,Olympus, sent an "URGENT-DEVICE CORRECTION " letter dated December 12, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were informed that any current irrigation tubing in their inventory can be used with the OFP pump; however, they must inspect the tubing installed on the OFP pump for signs of water leakage; if leakage is found, discontinue use of the tubing and replace with a new irrigation tube. In addition, complete and return the enclosed questionnaire via fax to (484) 896-7128. If you do not wish to use your current inventory of irrigation tubing and would like a replacement, call Olympus Customer Care Center at 1-800-848-9024. Olympus is discontinuing the current version of irrigation tubing and has redesigned the irrigation tubing to prevent water leakage. For any additional information, contact the V.P., Regulatory/Clinical Affairs & Quality Assurance at (484) 896-5688 or email at laura.storms-tyler@olympus.com.
- Initiated
- 2011-12-12
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-09-25
- Distribution
- Nationwide distribution: USA including states of: AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, AND WY.
- Recall number
- Z-0806-2012
- Recall status
- Terminated
- CDRH record ID
- 105979
- FDA classified
- 2012-01-18
- Recall event number
- 60613