Open original FDA record ↗Recall record
- FDA identifier
- Z-0806-2016
- Company in FDA record
- Viewray Incorporated
- FDA classification date
- Feb 12, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The system loaded a completion fraction in the incorrect order after a treatment interruption.
Source facts
- Action
- ViewRay sent an Customer Advisory Notice dated December 24, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Workaround: If you want to move a completion plan to a later date in the Delivery Calendar then do not schedule it until right before you are ready to treat that completion plan. While we will address this issue in the next release of software we recommend that in the event of a completion plan you always review the plan name loaded on the TPDS and verify the remaining time and beams are as expected with the completion plan. If you have any questions or concerns regarding this issue, please do not hesitate to contact ViewRay Customer Support at support@viewray.com or call the ViewRay" Customer Support center at 855-286- 8875.
- Initiated
- 2015-12-24
- Root cause
- Software design
- Terminated
- 2017-02-22
- Distribution
- US Distribution to the states of : CA, FL, MO & WI., and Internationally to Republic of Korea.
- Recall number
- Z-0806-2016
- Recall status
- Terminated
- CDRH record ID
- 143035
- FDA classified
- 2016-02-12
- Recall event number
- 73086