Z-0806-2020: Kodak DirectView DR3000/3500 - Product Usage: permanently installed as a diagnostic system intended to ge

Recall record Z-0806-2020 for Carestream Health, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0806-2020
Company in FDA record
Carestream Health, Inc.
FDA record creation date
Jan 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Carestream Health has discovered a potential safety problem that can lead to unintended movement of the U-Arm.

Source facts

Action
On January 6, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Users were informed that they may continue to use the system at this time. However, until the correction is executed, Carestream Health strongly recommends against performing any equipment positioning while the patient is near the system to avoid any bodily injury. If the user of the equipment observes any abnormal displacement in the vertical axis during the vertical movement, the device should be removed from use and a service call placed to Carestream (or local dealer) without delay. A Carestream Service Engineer or Carestream Health Authorized Representative will contact user sites and set up a service call to inspect user's devices within the next 6 months to determine whether the inverter parameters are set correctly. If adjustment is needed, the settings will be corrected and proper U-Arm operation will be verified. If you have any questions or concerns, please contact the Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis. Outside of the U.S., please call your local Carestream Service support number. If you have distributed the device outside your facility, please alert your customer(s) of this field correction and contact the Carestream Customer Care Center as instructed above.
Initiated
2020-01-06
Root cause
Component change control
Terminated
2020-05-29
Distribution
Worldwide distribution in the sate of Iowa and countries of Italy, China, Spain, and Austria.
Recall number
Z-0806-2020
Recall status
Terminated
CDRH record ID
178821
FDA record created
2020-01-13
Recall event number
84616

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗