Skip to content

Public FDA record

Z-0806-2025: EnVisio Patient Pad Transition Cushion, Model/Catalog Number: ENVISIO-SYS Product Description: The EnVisi

Recall record Z-0806-2025 for Elucent Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0806-2025
Company in FDA record
Elucent Medical Inc
FDA classification date
Jan 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation System.

Source facts

Action
An "Urgent Medical Device Recall" letter dated 11/18/24 was sent to customers. How to Recognize That the Device May Fail: An audible sound is made by the compression of the EnVisio Patient Pad Transition Cushion prior to the back section disassociation from the bed frame. We are aware of one product event associated with one complaint involving this issue. Component Which is the Subject of This Recall: The subject of this recall is the EnVisio Patient Pad Transition Cushion as it pertains to the inclining of Skytron Beds greater than 40 degrees. Actions to be Taken by the Customer: Due to potential harm resulting from the back section disassociating from the bed frame of Skytron beds, Elucent Medical is recommending not inclining Skytron beds beyond 40 degrees with the EnVisio Patient Pad Transition Cushion present. The Patient Pad Transition Cushion can be used safely when the bed itself is not inclined to such a degree. Elucent is working on an engineering solution that will allow increased range of incline of Skytron model beds. A replacement Patient Pad Transition Cushion will be provided to you as soon as possible. Actions to be Taken by Elucent Medical: Elucent Medical will replace all Patient Pad Transition Cushions as soon as a suitable replacement is tested and available. Contact Information Please address all specific questions regarding this recall to Elucent Customer Service via telephone (844-417-1700) [Monday Friday, 7:00am 5:00 pm Central Time] or email (customerservice@elucent.com).
Initiated
2024-11-18
Root cause
Device Design
Distribution
US Distribution: AZ, CA, CO, CT, FL, GA, IL, KY, MN, NC, ND, NJ, NY, OH, PA, SC, WI
Recall number
Z-0806-2025
Recall status
Open, Classified
CDRH record ID
211491
FDA classified
2025-01-06
Recall event number
95902

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗