A problem with software for the PolyTiter Immunofluorescent Titration System in that no flags or warnings appear when PolyTiter Calibrator Values are re-entered or changed (including values changed to unacceptable values) after the initial curve has been generated, if the 'refresh button' has not been pressed. The graph may appear valid; however erroneous results may occur.
Source facts
Action
Customers were sent a Technical Bulletin describing the correct procedure to be used when entering data into the Poly/Titer tiration system, until the updated software is installed
Initiated
2004-09-24
Root cause
Other
Terminated
2006-07-22
Distribution
OH, NJ, KY, TX, NC, FL, VA, WA, AL, CO, MO, AZ , CA , CT & NV
Recall number
Z-0807-05
Recall status
Terminated
CDRH record ID
36879
FDA classified
2005-05-10
Recall event number
30946