Z-0807-2022: BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL

Recall record Z-0807-2022 for Bio-Rad Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0807-2022
Company in FDA record
Bio-Rad Laboratories, Inc.
FDA record creation date
Mar 16, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet stability specifications

Source facts

Action
On 02/08/2022, the firm sent out an "UGENT: Medical Device Correction" letter to customers via overnight FedEx informing them that the Recalling Firm continues to receive reports of false reactive RPR results despite labeling revisions and the previous Technical Bulleting dated December 2021 (LB003034revA). This is in response to ongoing concerns related to COVID-19 vaccine interference with RPR and the stability issue that led to the short dating of lot, the Recalling Firm is taking the additional field action to disable the RPR portion of the Syphilis Total and RPR panels until further notice. Customers are instructed to: 1. Dispose of the reagents according to local waste management procedures. 2. If they have further distributed or transferred the product to other customers, or labs from their site, notify those customer of this field action. 3. Complete Customer Response Form and email or fax the complete response form to local Bio-Rad Technical Support or email or fax the complete response form to your local Bio-Rad Technical Support. Please return this form to Techsupport.ussd@bio-rad.com or the Toll Free Fax 1-888-228-0688 within 10 days of receiving this notice. within 10 days of receiving this notice. The firm will be disabling (turn off the RPR analyte in the software) the RPR assay beginning with the reagent (Lot No. 301399 and calibrator (Lot No. 53440). If customers laboratory use of this reagent and calibrator, they will be unable to report RPR results. Customers are provided with directions for turning off the RPR analyte (see specific instructions attached to Customer Notification Letter). For questions or assistance, contact Bio-Rad Technical Support at email or fax the complete response form to your local Bio-Rad Technical Support. Please return this form to Techsupport.ussd@bio-rad.com or the Toll Free Fax 1-888-228-0688
Initiated
2022-02-08
Root cause
Under Investigation by firm
Distribution
Worldwide distribution U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of France, Germany, Italy, Spain, Saudi Arabia, and United Arab Emirates.
Recall number
Z-0807-2022
Recall status
Open, Classified
CDRH record ID
192220
FDA record created
2022-03-16
Recall event number
89684

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗