Skip to content

Public FDA record

Z-0807-2025: 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENG

Recall record Z-0807-2025 for Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0807-2025
Company in FDA record
Stryker Corporation
FDA classification date
Jan 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.

Source facts

Action
On December 3, 2024, the firm notified affected consignees via "Urgent Medical Device Field Action" letters dated December 2, 2024. Customers were instructed to locate Neptune S Waste Management Systems and to quarantine affected products at their facility; to contact their Stryker Sales Rep to schedule the replacement of the affected device; to complete Business Reply Form after scheduling replacement and email completed form to instruments.recalls@stryker.com and maintain information internally until all required action have been completed. Please contact Recall Coordinator at 269-290-0524 or Instruments.recalls@stryker.com with questions or concerns. An URGENT Medical Device Field Action dated 10/13/25 was sent to customers. The purpose of this notification is to advise you that Stryker Instruments has expanded the scope of impacted lots for the Neptune S Waste Management System recall. Due to a distribution error, your facility may have received a recalled device. Next Steps: 1. Locate Neptune S Waste Management Systems and quarantine affected products at your facility Reference the catalog and affected serial numbers in the table above to know if your product is impacted. 2. Contact your Stryker Sales Rep. Your rep will help you schedule the repair of your Neptune device. 3. Complete Business Reply Form (BRF) Once you have scheduled your repair, fill out the attached BRF on page 2. Please email completed form to instruments.recalls@stryker.com. 4. Maintain this notification internally until all required actions have been completed. Please contact Brooke Thompson at 269-290-0524 or Instruments.Recalls@stryker.com with questions or concerns.
Initiated
2024-12-03
Root cause
Under Investigation by firm
Distribution
US (Domestic) distribution to states of: WA, GA, and IL. OUS (International) distribution to Japan and Netherlands.
Recall number
Z-0807-2025
Recall status
Open, Classified
CDRH record ID
211411
FDA classified
2025-01-06
Recall event number
95856

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗