Open original FDA record ↗Recall record
- FDA identifier
- Z-0808-05
- Company in FDA record
- Medtronic Perfusion Systems
- FDA classification date
- May 10, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some warehouse inventory of Custom Pack product bags were found with ruptured seals. The seals were partially opened in the center of the bag's Tyvek edge. Since this bag is used to enclose and seal the custom pack assembly & had been sterilized, the ruptured seal breaches the sterility barrier for the package.
Source facts
- Action
- US customers were sent certified letters beginning 03/25/05 identifying the situation, cautioning customers about the possibility of an open bag with loss of sterility and requested return of all product to Medtronic Perfusions Systems at 7611 Northland Blvd., Brooklyn Park, MN. Customers are also requested to complete a Field Corrective Action Certificate acknowledging the receipt of this information.
- Initiated
- 2005-03-25
- Root cause
- Other
- Terminated
- 2005-11-17
- Distribution
- nationwide, Canada, Australia, Japan and Latin America
- Recall number
- Z-0808-05
- Recall status
- Terminated
- CDRH record ID
- 37964
- FDA classified
- 2005-05-10
- Recall event number
- 31477