Z-0808-06: Ventilator-PLV Continuum (PLVC I), Model Number: P1000. The PLVC I is a microprocessor controlled, co

Recall record Z-0808-06 for Respironics California Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0808-06
Company in FDA record
Respironics California Inc
FDA classification date
May 2, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Flow valve failures- When failure occurs, air flow from the ventilator ceases resulting in the PLVC I transitioning into a 'Vent Inop' mode whereby the safety valve is activated, opening the circuit to ambient air. Ventilator dependent patients may not receive adequate ventilatory support if this problem occurs.

Source facts

Action
Notification letters were mailed to customers on March 20, 2006 notifying them of this action. Customers were instructed to quarantine all PLV Continuum Ventilators in their possession and not to place patients on them. Respironics Customers Service will contact customers to arrange for the return of all PLV Continuum Ventilators. Customers were instructed to safely plan and transition patients in their care from the PLV Continum Ventilator onto to other comparable patient support devices. If customers did not have a suitable ventilator to use for their patients, Respironics will loan them a PLV 100, 102 or 102b Ventilator with no charge. If patient support goes beyond the capabilities of the PLV 100, 102, or 102b Ventilator is required, Respironics will make arrangements to loan the customer with a suitable substitute rental ventilator at no charge. The field action will involve the upgrade of all affected devices with a new flow valve assembly. All US and International units are being recalled to the factory in Carlsbad, California. The rework will be conducted on-site.
Initiated
2006-03-20
Root cause
Other
Terminated
2008-04-21
Distribution
Nationwide, plus Australia, Argentina, Canada, Japan, Hong Kong, Netherlands, Saudi Arabia, & Taiwan
Recall number
Z-0808-06
Recall status
Terminated
CDRH record ID
45200
FDA classified
2006-05-02
Recall event number
35056

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗