Z-0809-2021: Disposable Humidification Flask, Reference Part Number K-MINC-CTS-S, Order Number G32707

Recall record Z-0809-2021 for Cook Medical Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0809-2021
Company in FDA record
Cook Medical Llc
FDA record creation date
Jan 8, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A small percentage of the Disposable Humidification Flasks may have a manufacturing defect that can cause the silicone stoppers to dislodge or gas to leak from the flasks, impeding gas flow to the incubation chambers of the Mini Incubator.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 12/16/20 was sent to customers. Action to be Taken: As a precautionary measure, we recommend that users perform two in-house checks prior to using a new flask to improve detection of the manufacturing defect. 1. Visual Inspection Check the flask assembly: when the stopper is placed on the flask, the tab on the silicone stopper should engage with the indent in the flask. Flasks that have the defect may not allow the stopper to be fully pushed into the indent of flasks, as shown in Figure 1. 2. Leak Test Check for leaks: Flasks that have the defect will cause a gas leak that will impede gas flow into the Mini Incubator. a. Prepare the Mini Incubator by removing any flasks. b. Fill the flask to be tested with sterile distilled water (to the fill line) and insert it into the incubator. c. Set the flow rate to 15mL/min and wait for the purge cycle to complete. d. Block the gas flow from both tubes exiting the flask by kinking and holding the tubes as shown in Figure 2. i. When the tubes are folded, the area around the folds may appear white this does not indicate that the silicone tubing has been damaged. e. If the gas flow does not fall to zero within 45 seconds, the flask may be defective. f. Unblock the tubes as soon as the flow reaches zero; blocking the tubes for more than two minutes may cause pressure to build up in the flask and the silicone stopper to suddenly pop off. i. Gas flow may be greater than 15mL/min immediately after the tubes are unblocked. This is normal the incubator will gradually adjust the gas flow rate to the correct value. g. If a potential defect is discovered, please take a photo and report this to Cook Medical Customer Relations, customerrelationsNA@cookmedical.com or 1-800-4574500, extension 152146, referencing FAIP 2020-0006 , in order to obtain a credit for the faulty product. Do not use the flask if a potential defect is discovered. Dispose of
Initiated
2020-12-16
Root cause
Under Investigation by firm
Terminated
2023-11-08
Distribution
US Nationwide
Recall number
Z-0809-2021
Recall status
Terminated
CDRH record ID
184946
FDA record created
2021-01-08
Recall event number
86965

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗