Open original FDA record ↗Recall record
- FDA identifier
- Z-0809-2025
- Company in FDA record
- Spacelabs Healthcare, Inc.
- FDA classification date
- Jan 7, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
Source facts
- Action
- On 11/25/2024, correction notices were sent to customers asking them to do the following: 1) Service manual has been updated by adding a section for "Required periodic system maintenance of Spacelabs Telemetry Farm", pages 5-30 through 5-35. 2) Download current versions of the Service Manual and PM Instructions at https://manuals.spacelabshealthcare.com/ 3) You may also obtain a copy of the latest manual by contacting the firm's Technical Support. 4) Discard previous copies of these documents 5) Complete the online customer acknowledgement form which can be accessed by using URL: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-windows-web-proxy-issue/ For additional information or to report any adverse events as a result of this issue, please contact the firm's Technical Support 1-800-522-7025 and select 2 for Technical Support
- Initiated
- 2024-11-25
- Root cause
- Labeling design
- Distribution
- US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic
- Recall number
- Z-0809-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211545
- FDA classified
- 2025-01-07
- Recall event number
- 95927