Skip to content

Public FDA record

Z-0809-2025: Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with

Recall record Z-0809-2025 for Spacelabs Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0809-2025
Company in FDA record
Spacelabs Healthcare, Inc.
FDA classification date
Jan 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.

Source facts

Action
On 11/25/2024, correction notices were sent to customers asking them to do the following: 1) Service manual has been updated by adding a section for "Required periodic system maintenance of Spacelabs Telemetry Farm", pages 5-30 through 5-35. 2) Download current versions of the Service Manual and PM Instructions at https://manuals.spacelabshealthcare.com/ 3) You may also obtain a copy of the latest manual by contacting the firm's Technical Support. 4) Discard previous copies of these documents 5) Complete the online customer acknowledgement form which can be accessed by using URL: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-windows-web-proxy-issue/ For additional information or to report any adverse events as a result of this issue, please contact the firm's Technical Support 1-800-522-7025 and select 2 for Technical Support
Initiated
2024-11-25
Root cause
Labeling design
Distribution
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic
Recall number
Z-0809-2025
Recall status
Open, Classified
CDRH record ID
211545
FDA classified
2025-01-07
Recall event number
95927

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗