Open original FDA record ↗Recall record
- FDA identifier
- Z-0811-2010
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Feb 22, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Affected devices contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock. There could be a delay in therapy or prevention of defibrillation therapy if the defibrillator were needed in a cardiac arrest situation. Delay or prevention of defibrillation therapy may contribute to the death of a viable patient.
Source facts
- Action
- Consignees were sent a notification letter dated March 2009 that explained firm representatives would follow up with each customer within 30 days to arrange an update to affected devices. Affected Biphasic PCBAs will be replaced. Customers advised to keep the devices in use and to continue to perform daily User Test.
- Initiated
- 2009-03-31
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-08-08
- Distribution
- Distributed domestically and internationally to hospitals, clinics, etc. Foreign distribution to the following countries: Australia, Brazil, Canada, Colombia, Costa Rica, Germany, Hong Kong, Mexico, Netherlands, and Puerto Rico.
- Recall number
- Z-0811-2010
- Recall status
- Terminated
- CDRH record ID
- 87123
- FDA classified
- 2010-02-22
- Recall event number
- 54005