Open original FDA record ↗Recall record
- FDA identifier
- Z-0811-2014
- Company in FDA record
- Ziemer Usa Inc
- FDA classification date
- Jan 23, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A defect in the controller board can lead to the condenser to the video monitor becoming too warm. This could potentially cause the video monitor to shut off.
Source facts
- Action
- A recall notification letter dated April 16, 2013 was mailed to direct accounts. The letter notified the direct accounts of the problem and advised them that a technician would be arriving at their facility during the week of June 2, 2013 to replace the defective controller board. The technician will also perform preventative maintenance on the device at the same time. Direct accounts are advised to contact Ziemer USA, Inc. at 618-462-9301 if they have any questions about this issue.
- Initiated
- 2013-04-16
- Root cause
- Component design/selection
- Terminated
- 2017-12-27
- Distribution
- Worldwide Distribution: US (nationwide) including states of: NJ and IL; and Internationally to: Finland, Japan, and Germany.
- Recall number
- Z-0811-2014
- Recall status
- Terminated
- CDRH record ID
- 117554
- FDA classified
- 2014-01-23
- Recall event number
- 64966