Skip to content

Public FDA record

Z-0811-2018: NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UD

Recall record Z-0811-2018 for NxStage Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0811-2018
Company in FDA record
NxStage Medical, Inc.
FDA record creation date
Mar 1, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard dialysis fluid of 0.5 EU/ml when used to prepare product.

Source facts

Action
On January 29, 2018 Product Recall Notices & Recall Reply Forms included were emailed to those patients for whom the firm had an email address. Hard copy letters were sent via UPS second day air to patients for whom NxStage did not have an email address. All Dialysis centers were contacted via email. In addition to the active patients and dialysis center consignees of affected product, recall letters were also emailed to the Dialysis Center Home Training Nurse Manager for all active or dropped patients who were shipped affected product, and corporate contacts for national dialysis accounts. Customers are advised to inspect inventory, quarantine, discontinue use and return product. Customers are also encouraged to return the completed Recall Reply Form. Contact NxStage Customer Service to arrange for return of all affected product and for replacement product to be sent. Customers with questions may call Customer Service at 1-866-697-8243.
Initiated
2018-01-29
Root cause
Material/Component Contamination
Terminated
2020-04-03
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : United Kingdom & Spain
Recall number
Z-0811-2018
Recall status
Terminated
CDRH record ID
161554
FDA record created
2018-03-01
Recall event number
79185

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗