Z-0812-04: Medtronic GEM DR Implantable Cardioverter Defibrillators, Model 7271

Recall record Z-0812-04 for Medtronic Inc. Cardiac Rhythm Managment. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0812-04
Company in FDA record
Medtronic Inc. Cardiac Rhythm Managment
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some of the devices with suspect capacitors (high voltage capacitors associated with prior notifications in 1999 and in 2000) have had unexpected charge circuit time-outs or charge circuit inactive conditions as the battery voltage nears the level for Elective Replacement of the devices.

Source facts

Action
A letter, dated April 2004, was sent on April 5, 2004 to physicians following patients who had the defibrillators implanted. The letter gave recommendations to physicians for selective replacement of some of the implanted devices and for avoiding the problem. Physicians are being visited by sales representatives to verify notification.
Initiated
2004-04-02
Root cause
Other
Terminated
2005-11-29
Distribution
Nationwide in the United States and worldwide.
Recall number
Z-0812-04
Recall status
Terminated
CDRH record ID
32530
FDA classified
2004-07-20
Recall event number
28798

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗