Skip to content

Public FDA record

Z-0812-2018: 731 Series Ventilators running software version 05.20.00 The devices in the ZOLL ventilator are indicat

Recall record Z-0812-2018 for ZOLL Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0812-2018
Company in FDA record
ZOLL Medical Corporation
FDA record creation date
Mar 1, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software anomaly in the 731 software version 05/20/00, was identified, which can lead to a user inadvertently changing device settings.

Source facts

Action
On June 30, 2017 an urgent recall letter was mailed to customers with the following instructions: Customers who have affected devices should immediately take the following steps: (1) Alert all users of 731 Ventilator models to this problem. (2) Direct users to always verify device parameters after selecting the device Confirm button. (3) Contact ZOLL's Technical Service Department or your local ZOLL Service Provider to schedule a software update for affected devices Our 24/7 technical support numbers 1(800) 348-9011 or +l (978) 421- 9460 are available to assist users with any aspect of this notice.
Initiated
2017-06-30
Root cause
Under Investigation by firm
Terminated
2018-10-09
Distribution
Outside US
Recall number
Z-0812-2018
Recall status
Terminated
CDRH record ID
161746
FDA record created
2018-03-01
Recall event number
79248

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗