A software anomaly in the 731 software version 05/20/00, was identified, which can lead to a user inadvertently changing device settings.
Source facts
Action
On June 30, 2017 an urgent recall letter was mailed to customers with the following instructions:
Customers who have affected devices should immediately take the following steps: (1) Alert all users of 731 Ventilator models to this problem.
(2) Direct users to always verify device parameters after selecting the device Confirm button.
(3) Contact ZOLL's Technical Service Department or your local ZOLL Service Provider to schedule a software update for affected devices
Our 24/7 technical support numbers 1(800) 348-9011 or +l (978) 421- 9460 are available to assist users with any aspect of this notice.
Initiated
2017-06-30
Root cause
Under Investigation by firm
Terminated
2018-10-09
Distribution
Outside US
Recall number
Z-0812-2018
Recall status
Terminated
CDRH record ID
161746
FDA record created
2018-03-01
Recall event number
79248