The recommended hydration fluid for the 8180 Gram-Positive Blood Culture Control Panel (Inactivated Pellet), has changed from sterile 0.85% saline or sterile blood culture media to sterile DI water or pH 7.2 phosphate buffer.
Source facts
Action
Recall Letters was sent via email to all consignees regarding the correction on 12/16/2019 informing users of a new recommended hydration fluid which is reflected in labeling QC Sets and Panels: Technical Information TIB.2034 Rev G.
Questions can be directed at Technical Support team at 1.320.229.7064 or techsupport@microbiologics.com . A response form is requested to be returned.
Initiated
2019-12-16
Root cause
Device Design
Terminated
2020-12-11
Distribution
Distributed nationwide including Washington DC, and distributed internationally to Germany and Netherlands.
Recall number
Z-0812-2020
Recall status
Terminated
CDRH record ID
178514
FDA record created
2020-01-16
Recall event number
84524
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.