Labeling discrepancy: The manual instruction state that the line should be flushed with 5 mL of saline and the first 5 mL of blood or six dead space volumes should be discarded. The Cartridge Test Information sheets that it should be 2 mL of blood - which is incorrect.
Source facts
Action
Abbott Point of Care issued an "Important Product Information" letter beginning August 18, 2009. Consignees were instructed on how to reconcile all affected product.
For further information, contact Abbott Point of Care Technical Support at 1-800-366-8020, Option 1.
Initiated
2009-08-18
Root cause
Labeling False and Misleading
Terminated
2010-05-18
Distribution
Worldwide distribution.
Recall number
Z-0813-2010
Recall status
Terminated
CDRH record ID
83881
FDA classified
2010-02-23
Recall event number
52754