Open original FDA record ↗Recall record
- FDA identifier
- Z-0813-2018
- Company in FDA record
- Alto Development Corp
- FDA record creation date
- Mar 1, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Surgical punches are being recalled due to potential packaging damage which may have occurred during transportation resulting in loss of sterility.
Source facts
- Action
- A&E Medical sent an Recall Notice Action Required letter dated October 20, 2017. Surgical Punches are being recalled due to potential packaging damage which may have occurred during transportation resulting in a loss of sterility. Recent testing identified certain shipping configurations which may cause damage to the sterile barrier pouch if the Surgical Punches experience excessive movement from multiple drops during transport A&E is recalling all product currently on the market. Customer and Distributor Letters were sent with Acknowledgement Form. Product will be quarantined pending rework. For further questions, please call (800) 323-4035.
- Initiated
- 2017-10-23
- Root cause
- Package design/selection
- Terminated
- 2019-04-30
- Distribution
- Worldwide Distribution - USA Distribution
- Recall number
- Z-0813-2018
- Recall status
- Terminated
- CDRH record ID
- 161805
- FDA record created
- 2018-03-01
- Recall event number
- 79274