Z-0813-2019: Naera Hospital Bassinet, Model No. 4402

Recall record Z-0813-2019 for Stryker Medical Division of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0813-2019
Company in FDA record
Stryker Medical Division of Stryker Corporation
FDA classification date
Feb 7, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was identified that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub. A Health Hazard Evaluation was completed which identified the potential risk of patients moving out of position and rolling over on the mattress. While the highest possible severity of harm associated with this condition is death, there have been no reports of injury or harm. This condition has only presented in regions of high altitude; however, Stryker has chosen to replace all Naera mattresses in the field.

Source facts

Action
The firm notified customers on October 10, 2018, via Urgent Medical Device Recall letters. The letters informed users of the issue with the mattress component of the bassinet. Customers were instructed to do the following: 1. Find the serial numbers listed on the attached business reply form and be sure you know where they are located at your facility. 2. Return the enclosed business reply form to confirm receipt of this notification by fax 269 488 8691 or email productfieldaction@stryker.com. 3. Inspect the Nra bassinets at your location. If any exhibit the condition as shown in the Product issue section of this letter, please quarantine the unit, remove it from service, and contact Stryker at 269 389 8306. 4. One of Strykers representatives will contact you to schedule the replacement of the mattresses in all affected units at your location. 5. If you have questions regarding the continued safe use of your product, please contact Technical Support by calling +1 800 327 0770 and selecting opt 2. 6. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location in the space provided on the business reply form. 7. If you have disposed of any of these units and they are no longer in use, please, advise us of their obsolescence by providing us with their serial number in the space provided on the business reply form. If you have any questions or concerns, please contact Customer Service +1 800 327 0770. The firm's normal business hours are Monday-Friday 8 a.m-6 p.m. (EST).
Initiated
2018-10-10
Root cause
Device Design
Terminated
2020-04-29
Distribution
Nationwide distribution. Foreign distribution to Argentina, Australia, Canada, Chile, China, Colombia, Germany, India, Japan, Netherlands, South Africa, Turkey, Saudi Arabia, United Kingdom, and Mexico.
Recall number
Z-0813-2019
Recall status
Terminated
CDRH record ID
168452
FDA classified
2019-02-07
Recall event number
81393

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗