Z-0813-2020: Carestation 620/650/650c A1 Product Usage: anesthesia systems are intended to provide general inhalation

Recall record Z-0813-2020 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0813-2020
Company in FDA record
GE Healthcare, LLC
FDA classification date
Jan 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high priority audio and visual alarms. Loss of mechanical ventilation could lead to hypoxia if the clinician does not intervene.

Source facts

Action
The firm initiated their field notification by letter on November 25, 2019. The letter states the reason for recall and provides the following safety instructions: You can continue to use the anesthesia system. - If you observe the message Ventilate Manually! , change from mechanical to manual ventilation. At any time, the clinician may use a self-inflating bag to ventilate the patient and/or switch to another anesthesia device. Contact your GE Healthcare representative for repair of the device. - Perform the planned maintenance (PM) every 12-months at a minimum per the User s Reference Manual which includes inspection of the cable connection. Note: This inspection step is included in the annual PM described in the Technical Reference Manual. Performing this step in the PM would confirm the integrity of the cable connection. Contact Information: If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. The firm will be correcting the units. The firm issued a Press Release on January 13, 2020.
Initiated
2019-11-25
Root cause
Process control
Terminated
2023-11-30
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Albania, Algeria, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia and Hrvatska, Czech Republic, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam.
Recall number
Z-0813-2020
Recall status
Terminated
CDRH record ID
178322
FDA classified
2020-01-23
Recall event number
84438

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗