Z-0813-2022: Outset Tablo Console Model Number: PN-0003000 Model Number: PN-0006000

Recall record Z-0813-2022 for Outset Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0813-2022
Company in FDA record
Outset Medical, Inc.
FDA record creation date
Mar 21, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.

Source facts

Action
On 01/20/2022, the firm send a "MEDICAL DEVICE FIELD ACTION" Letter to customers informing them that Outset Medical is proactively replacing a component on the Talbo Hemodialysis consoles due to the possibility of heat-related damage that may occur to the Tablo device. Outset Medical has identified that Tablo Hemodialysis Systems manufactured between July 6, 2018 and November 24, 2021 require the component replacement. Outset is indicating that the replacement will be completed within the next twelve months during the next routine maintenance or at the next service call. Outset Field Service Engineers will coordinate completion of the replacement. Customers are instructed to complete the acknowledgement form and promptly return to Outset Medical For questions, contact Outset Customer Experience Team at 844-698-2256.
Initiated
2022-01-20
Root cause
Device Design
Distribution
U.S. Nationwide distribution in the states of AL, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, MO, MT, NE, NV, NJ, NY, NM, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY.
Recall number
Z-0813-2022
Recall status
Open, Classified
CDRH record ID
192066
FDA record created
2022-03-21
Recall event number
89616

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗