BacT/VIEW will be unable to run scheduled macro applications unless the user touches the BacT/VIEW touch screen or clicks a function after resynchronizing the clocks.
Source facts
Action
A customer notification was sent on/about December 19, 2006. Customers will be sent a new BacT/VIEW software version C.30b QTR 1 2007.
Initiated
2006-12-19
Root cause
Other
Terminated
2009-12-08
Distribution
Worldwide, including USA, Canada, UK, Italy, France, Spain, Germany, Poland, Chile, Japan and the Netherlands.
Recall number
Z-0814-2007
Recall status
Terminated
CDRH record ID
49959
FDA classified
2007-05-18
Recall event number
37136