Skip to content

Public FDA record

Z-0814-2014: MRI Low Profile Single Lumen port with Open-Ended 6.6Fr Silicone Catheter. Product Code 0603880. Product

Recall record Z-0814-2014 for Bard Access Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0814-2014
Company in FDA record
Bard Access Systems
FDA classification date
Jan 23, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bard Access Systems is recalling various Low Profile Port with 6.6 Fr Single Lumen Broviac Catheter because a small number of Silicone catheters in the may be undersized to 5.3 Fr instead of 6.6 Fr as stated in the labeling for the device.

Source facts

Action
BARD sent an Urgent Product Recall letter dated October 2013 to affected customers via certified letter. The letter identified the affected product, problem and actions to be taken. Customers were asked to check all inventory locations, remove and return any unused affected prodcut to Bard Access Systems through return goods authorization number RAP80498 for a credit or replacement, and to complete the Reply Form and Inventory Reconcilation Form. For questions contact BAS Customer Service at 1-800-290-1689.
Initiated
2013-12-06
Root cause
Process control
Terminated
2014-08-05
Distribution
Worldwide Distribution - US nationwide, Belgium, and Canada.
Recall number
Z-0814-2014
Recall status
Terminated
CDRH record ID
124441
FDA classified
2014-01-23
Recall event number
66991

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗