Skip to content

Public FDA record

Z-0814-2018: OTTO BOCK Kenevo knee joints Model 3C60

Recall record Z-0814-2018 for Otto Bock Healthcare Product. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0814-2018
Company in FDA record
Otto Bock Healthcare Product
FDA record creation date
Mar 1, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.

Source facts

Action
On September 14, 2017 a letter was sent to all their consignees via 2nd day signature receipt required to customers instructing them to contact their patients and send in their devices for service within the first 2.3 years of use. The pylon clamp will be replaced at no charge. The pylon will also be inspected and replaced if necessary. The customer will be requested to complete a reply form indicating they have received the notice and intend to contact their patients to arrange for the requested service. Customer Service: (800) 328-3458 Option 1 Loaner Service: (800) 328-3458 Option 3, then Option 1
Initiated
2017-09-12
Root cause
Device Design
Terminated
2021-01-19
Distribution
AK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU
Recall number
Z-0814-2018
Recall status
Terminated
CDRH record ID
161205
FDA record created
2018-03-01
Recall event number
79019

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗