Skip to content

Public FDA record

Z-0814-2024: TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal secti

Recall record Z-0814-2024 for St. Jude Medical, Atrial Fibrillation Division, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0814-2024
Company in FDA record
St. Jude Medical, Atrial Fibrillation Division, Inc.
FDA classification date
Jan 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.

Source facts

Action
The recalling firm issued letters dated 12/18/2023 via hand-delivery and FedEx beginning 12/18/2023. The letter explained the issue, the scope of the problem, the impact and associated risks, and next steps to help reduce risk. These steps were: (1) Do not use any remaining inventory from the affected lots listed in Appendix A; (2) Complete and return the accompanying Acknowledgment Form; (3) Return all remaining unused devices from the affected lots; and (4) Forward the letter to anyone within their organization who may need to be notified. Replacements of returned affected devices with a similar curve are available and the customer's Abbott representative can assist in returning these devices and obtaining replacements.
Initiated
2023-12-18
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Recall number
Z-0814-2024
Recall status
Open, Classified
CDRH record ID
205276
FDA classified
2024-01-25
Recall event number
93711

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗