Z-0814-2025: Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1107-9604

Recall record Z-0814-2025 for GE Medical Systems, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0814-2025
Company in FDA record
GE Medical Systems, LLC
FDA classification date
Jan 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 11/18/24 was sent to customers. Actions to be taken by Customer/User You can continue to use the vaporizer in accordance with the instructions in the User Reference Manual (URM) and the actions described below: 1. Perform the Output Concentration Check described in Appendix 1 for every impacted vaporizer (see Affected Product Details in this letter). " If a vaporizer passes the Output Concentration Check, you can continue to use that vaporizer. GE HealthCare recommends use of anesthetic agent monitoring during vaporizer use. " If a vaporizer fails the Output Concentration Check, you can continue to use that vaporizer with anesthetic agent monitoring until GE HealthCare corrects the vaporizer. " If a vaporizer fails the Output Concentration Check and anesthetic agent monitoring is not available, remove the vaporizer from service. 2. Please complete and return the attached Medical Device Notification Acknowledgement Response Form to ARC.FMI34139@gehealthcare.com 3. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions. 4. Please retain this document for your records. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2024-11-18
Root cause
Process control
Distribution
US Nationwide. Global Distribution.
Recall number
Z-0814-2025
Recall status
Open, Classified
CDRH record ID
211401
FDA classified
2025-01-07
Recall event number
95855

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗