Skip to content

Public FDA record

Z-0815-2016: CEA / CD66e Ab-2, Product code RB-368-R7, RB-368-R1; 7 ml each. Description Carcinoembryonic antigen (

Recall record Z-0815-2016 for Lab Vision Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0815-2016
Company in FDA record
Lab Vision Corporation
FDA classification date
Feb 17, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Certain lots of CEA/CD66e Ab2 were produced with a dilution factor of 1:100 instead of the recommended 1:250.

Source facts

Action
ThermoFisher sent an Urgent Medical Device recall letter dated January 18, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed if they have any of the affected product to contact Lab Vision with the information requested on the attached Acknowledgment Form. Upon identification of the affected lot, please notify Lab Vision of the quantity previously used and the quantity still in inventory. If any of the above listed lots were previously used, please review test results to ensure a valid result was obtained. If a valid result was obtained, no further action is required. If high background staining was seen, standard IHC troubleshooting guidance for high background staining can be used. (See RISK TO HEALTH section of the letter.) Affected product may be returned to Lab Vision. Please indicate on the Acknowledgement Form whether the product will be returned. Upon receipt of the form, RGA information will be sent for product requiring return. If it is preferred to destroy the product at the customer site, please indicate this on the Acknowledgement form. Please use the attached Recall Return Response Acknowledgement & Receipt Form. Customers with questions were instructed to call 269-544-5628.
Initiated
2016-01-18
Root cause
Nonconforming Material/Component
Terminated
2016-06-24
Distribution
Worldwide Distribution - US including CA, NY, GA, MS, PA, MA, TX, KY, AZ. and Internationally to Greece, France, Germany, Singapore and Serbia.
Recall number
Z-0815-2016
Recall status
Terminated
CDRH record ID
143095
FDA classified
2016-02-17
Recall event number
73120

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗