Open original FDA record ↗Recall record
- FDA identifier
- Z-0815-2016
- Company in FDA record
- Lab Vision Corporation
- FDA classification date
- Feb 17, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Certain lots of CEA/CD66e Ab2 were produced with a dilution factor of 1:100 instead of the recommended 1:250.
Source facts
- Action
- ThermoFisher sent an Urgent Medical Device recall letter dated January 18, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed if they have any of the affected product to contact Lab Vision with the information requested on the attached Acknowledgment Form. Upon identification of the affected lot, please notify Lab Vision of the quantity previously used and the quantity still in inventory. If any of the above listed lots were previously used, please review test results to ensure a valid result was obtained. If a valid result was obtained, no further action is required. If high background staining was seen, standard IHC troubleshooting guidance for high background staining can be used. (See RISK TO HEALTH section of the letter.) Affected product may be returned to Lab Vision. Please indicate on the Acknowledgement Form whether the product will be returned. Upon receipt of the form, RGA information will be sent for product requiring return. If it is preferred to destroy the product at the customer site, please indicate this on the Acknowledgement form. Please use the attached Recall Return Response Acknowledgement & Receipt Form. Customers with questions were instructed to call 269-544-5628.
- Initiated
- 2016-01-18
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-06-24
- Distribution
- Worldwide Distribution - US including CA, NY, GA, MS, PA, MA, TX, KY, AZ. and Internationally to Greece, France, Germany, Singapore and Serbia.
- Recall number
- Z-0815-2016
- Recall status
- Terminated
- CDRH record ID
- 143095
- FDA classified
- 2016-02-17
- Recall event number
- 73120