The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator.
Source facts
Action
Del Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Initiated
2005-03-10
Root cause
Other
Terminated
2005-11-14
Distribution
Nationwide, Mexico and Thailand.
Recall number
Z-0816-05
Recall status
Terminated
CDRH record ID
37928
FDA classified
2005-05-11
Recall event number
31456