There may be a material defect in the up/down movement axle of the unit. This may cause the upper part of the unit to suddenly come down.
Source facts
Action
A recall letter, dated 04/28/2006, was sent to affected customers . The letter describes the issue, and requests that no patient operation is permitted until the unit is checked by authorized service technician. If the service technician noted wear of the screw, the assembly was replaced.
Initiated
2006-04-28
Root cause
Other
Terminated
2008-02-24
Distribution
Worldwide, including USA, Canada, Bermuda, and Colombia.
Recall number
Z-0816-2007
Recall status
Terminated
CDRH record ID
51835
FDA classified
2007-05-18
Recall event number
37838