Z-0816-2012: Sechrist Hyperbaric Chamber Gurney. The Sechrist Hyperbaric Chamber Gurney is an accessory device to the

Recall record Z-0816-2012 for Sechrist Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0816-2012
Company in FDA record
Sechrist Industries Inc
FDA classification date
Jan 19, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Stryker Medical has received complaints from customers that indicate a bushing in the braking system may fracture, which could result in the brake/steer pedal seizing. If the pedal seizes, the braking system may become inoperable or steering around corners may come more difficult. While no injuries have been associated with these reported events nor has there been reported failure events resulting

Source facts

Action
Sechrist Industries sent a Urgent Medical Device Correction letter dated September 9, 2009, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Locate the gurneys with the serial numbers listed in this advisory. 2. If any of the affected gurneys cannot be located, determine if they have been sold, transferred, or donated to another facility and notify us regarding their new location. 3.Verify the braking system is working properly. If the brake system is inoperable, or if the brake and /or steer require a greater than normal force to engage, remove the stretcher from service and contact Stryker Medical at 800-Stryker, option 3. Reference RA-2009-135. Do not put the stretcher back into service until is repaired by the Stryker field representative. 4. Contact Stryker Medical to schedule a service call for replacement of the bushing and mating parts of the stretcher. 5.Complete retrofit of the Gurney. 6.Verify correct function of the brake assembly following retrofit. 7.Complete and send the Correction Verification Form contained with this notice to Sechrist. For further questions please call (714) 579-8400. .
Initiated
2009-09-15
Root cause
Nonconforming Material/Component
Terminated
2012-01-19
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Peru, Venezuela, Brazil, Egypt, Mexico, Austrailia, Denmark and Taiwan
Recall number
Z-0816-2012
Recall status
Terminated
CDRH record ID
105498
FDA classified
2012-01-19
Recall event number
60466

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗