Open original FDA record ↗Recall record
- FDA identifier
- Z-0817-2011
- Company in FDA record
- DePuy Spine, Inc.
- FDA classification date
- Dec 27, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The scale which serves as a secondary means of identifying screw length is incorrectly etched on the screw caddy
Source facts
- Action
- Depuy Spine issued an "URGENT-VOLUNTARY PRODUCT RECALL NOTIFICATION" via e-mail on October 2, 2010 to customers. The notification describes the product, problem and action to be taken by customers. The customers were instructed to please distribute this information to appropriate personnel at their facility and return any caddies that they may have to their local distributor for DePuy Spine products. Depuy will return devices to customer promptly (within two business days), after reworking the devices by removing the scale altogether. Once the product is fully corrected, Depuy will provide the customers with a new and updated unit with the corrected scale markings. If you should have any questions or concerns regarding this action, please contact DePuy Spine Customers Service at 800-227-6633.
- Initiated
- 2010-10-02
- Root cause
- Labeling design
- Terminated
- 2011-04-15
- Distribution
- Nationwide distribution: USA including states of: AR, AZ, CA, FL, HI, IA, IL, KY, LA, MA, MI, MN, MO, NC, NY OH, PA, TN, TX, VA, and WA.
- Recall number
- Z-0817-2011
- Recall status
- Terminated
- CDRH record ID
- 94877
- FDA classified
- 2010-12-27
- Recall event number
- 56894