Open original FDA record ↗Recall record
- FDA identifier
- Z-0817-2012
- Company in FDA record
- Hospira Inc.
- FDA classification date
- Jan 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Silicone applied to the piercing pin of the blood set is reducing retention force of the piercing pin while in use. This could cause the piercing pin to slip out of the administration container.
Source facts
- Action
- The firm, Hospira, sent an "URGENT DEVICE RECALL" letter dated January 12, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory and immediately quarantine any affected product. Additionally, a Reply Form was attached for customers to complete and return via fax to 1-866-912-8410 or e-mail to: hospira2573@stericycle.com. Please contact Hospira Customer Care at 1-877-946-7747 for questions regarding this recall.
- Initiated
- 2012-01-12
- Root cause
- Component design/selection
- Terminated
- 2016-08-29
- Distribution
- Distributed in the states of North Carolina and Washington.
- Recall number
- Z-0817-2012
- Recall status
- Terminated
- CDRH record ID
- 106687
- FDA classified
- 2012-01-19
- Recall event number
- 60869