Skip to content

Public FDA record

Z-0817-2012: Hospira Blood Set, with 200 Micron Filter, 80 Inch, Non-DEHP; an Rx sterile medical device used to admini

Recall record Z-0817-2012 for Hospira Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0817-2012
Company in FDA record
Hospira Inc.
FDA classification date
Jan 19, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Silicone applied to the piercing pin of the blood set is reducing retention force of the piercing pin while in use. This could cause the piercing pin to slip out of the administration container.

Source facts

Action
The firm, Hospira, sent an "URGENT DEVICE RECALL" letter dated January 12, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory and immediately quarantine any affected product. Additionally, a Reply Form was attached for customers to complete and return via fax to 1-866-912-8410 or e-mail to: hospira2573@stericycle.com. Please contact Hospira Customer Care at 1-877-946-7747 for questions regarding this recall.
Initiated
2012-01-12
Root cause
Component design/selection
Terminated
2016-08-29
Distribution
Distributed in the states of North Carolina and Washington.
Recall number
Z-0817-2012
Recall status
Terminated
CDRH record ID
106687
FDA classified
2012-01-19
Recall event number
60869

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.