When using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual.
Source facts
Action
Customers were notified by letter on March 23, 2004 to contact their LIS vendor and to reconfigure the ASTM Field 10.1.4 to allow a maximum character length of 20 characters.
Initiated
2004-03-19
Root cause
Other
Terminated
2005-11-03
Distribution
Nationwide and to the following foreign countries: Chile, Canada, Germany, Taiwan, Australia, Signapore
Recall number
Z-0818-04
Recall status
Terminated
CDRH record ID
32420
FDA classified
2004-07-20
Recall event number
28728
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.