- FDA identifier
- Z-0819-06
- Company in FDA record
- Alpha Omega Services Inc
- FDA classification date
- Apr 27, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is a possibility that the diameter of the catheter does not meet specifications. As a result it may not be possible to correctly connect the catheter to a VariSource Quick Connect. This would prevent any treatment. Furthermore, since the problem would not be apparent until after the catheter has been implanted an unnecessary surgical implantation procedure would have taken place.
Source facts
- Action
- Recall Notification Letter: began February 22, 2006. Depth/Scope: 100% of the customer''s who purchased the affected products will be identified and contacted with a faxed and mailed Medical Device Recall notice and followed up with a phone call by the AOS Sales & Marketing personnel. All customers will be requested return all affected products, both used and unused.
- Initiated
- 2006-02-21
- Root cause
- Other
- Terminated
- 2006-08-30
- Distribution
- Nationwide and Argentina, Brasil, Canada, Israel, Korea, , Taiwan, United Kingdom & Venezuela
- Recall number
- Z-0819-06
- Recall status
- Terminated
- CDRH record ID
- 45045
- FDA classified
- 2006-04-27
- Recall event number
- 34969