Open original FDA record ↗Recall record
- FDA identifier
- Z-0819-2010
- Company in FDA record
- Veridex, LLC
- FDA classification date
- Feb 23, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Veridex has received 4 complaints regarding reagent bottles having been incorrectly placed or duplicate bottles places within the reagent tray of CellSearch Circulating Tumor Cell Kit. Delayed detection of disease relapse or progression may occur due to false low CTC results.
Source facts
- Action
- Urgent Product Correction Notification letters were sent to all customers on August 12, 2009 by Federal Express. Customers were advised to inspect the kits to assure the proper reagents are in the proper position. If not, do not use the kits and contact the company at 877-837-4339 (chose option 1).
- Initiated
- 2009-08-12
- Root cause
- Other
- Terminated
- 2012-07-18
- Distribution
- Product was distributed worldwide, US (one government account: National Cancer Institute, Frederick, MD). Also 8 international accounts in Belgium, France, Germany, Italy, Spain, United Kingdom, Netherlands, Norway and Israel.
- Recall number
- Z-0819-2010
- Recall status
- Terminated
- CDRH record ID
- 84478
- FDA classified
- 2010-02-23
- Recall event number
- 53025