Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
Source facts
Action
The firm initiated their correction on 12/16/2019 by letter. The letter explained the issue and directed the consignee to follow instructions in the letter until an update to the Centricity Universal Viewer can be made by the firm to correct the issue.
Initiated
2019-12-16
Root cause
Software design
Terminated
2023-10-05
Distribution
Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
Recall number
Z-0819-2020
Recall status
Terminated
CDRH record ID
178662
FDA classified
2020-01-17
Recall event number
84584