Open original FDA record ↗Recall record
- FDA identifier
- Z-0820-2007
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- May 26, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Quality control and patient results can be lower than expected when the true Troponin I value is less than or equal to 0.2 ng/mL which may result in the device failing to detect myocardial injury.
Source facts
- Action
- On 4/27/07, Ortho-Clinical Diagnostics in Rochester, NY, notified the end users of the recall of the product by phone, fax, and FedEx overnight letter. Clinical labs in possession of these lots have been contacted by OCD, and were instructed to (a) discontinue use of this product and (b) notify healthcare providers who ordered the test in recent weeks. Replacement product has been provided to clinical labs with the affected product lots. Clinical laboratories with questions may contact the company at 1-800-421-3311. On 4/27/07, OCD foreign affiliates were informed of this action by e-mail and instructed to inform their customers who received these reagent packs of this issue. Ortho-Clinical Diagnostics in Raritan, NJ, issued a nationwide press release on 5/04/07.
- Initiated
- 2007-04-27
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-09-07
- Distribution
- Worldwide-including USA, Canada, and Bermuda
- Recall number
- Z-0820-2007
- Recall status
- Terminated
- CDRH record ID
- 52013
- FDA classified
- 2007-05-26
- Recall event number
- 37873