Skip to content

Public FDA record

Z-0821-2007: Definium AMX 700 Digital mobile x-ray system, GE Healthcare 3000 North Grandview Waukesha, WI

Recall record Z-0821-2007 for GE Medical Systems LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0821-2007
Company in FDA record
GE Medical Systems LLC
FDA classification date
May 22, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

When using the Tech Switch on exposures >750ms, >100 kVp and > 100 mAs the exposure terminates prematurely.

Source facts

Action
Visit to each consignee beginning December 21, 2006.
Initiated
2006-12-21
Root cause
Other
Terminated
2008-02-24
Distribution
Product was distributed in the following states: MS, NE, NJ, and, TX
Recall number
Z-0821-2007
Recall status
Terminated
CDRH record ID
51936
FDA classified
2007-05-22
Recall event number
37848

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.