Z-0821-2015: VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1
Recall record Z-0821-2015 for MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK. View source facts, labeled dates, and the original FDA record.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
FDA classification date
Dec 17, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result, ExacTrac may display a Digitally Reconstructed Radiograph (DRR) image taken from the angle for the 1st port image fusion, rather than the one taken from the angle for the intended port
Source facts
Action
Mitsubishi sent a Field Safety Notice dated November 14, 2014, that was e-mailed to US end users on December 2, 2014, which provided actions to be taken by the end users. If users need any further information or support concerning this issue, they may contact MHI's Customer Service Representative, Brainlab AG in Germany, contact person: Franz Gum, Tel: +49-89-99-1568-0, Fax: +49-89-99-1568-33, E-mail: Franz.Gum@brainlab.com.
If any other abnormality is detected, stop operation and contact MHI's Customer Service representative immediately. --- An Updated Field Safety Notice (dated December 4, 2014,) was sent to end users to correct some terms used in the initial notification.
Initiated
2014-12-02
Root cause
Software design
Terminated
2017-06-26
Distribution
US Distribution to the states of NY, FL, TX and OH.
Recall number
Z-0821-2015
Recall status
Terminated
CDRH record ID
131960
FDA classified
2014-12-17
Recall event number
69825