There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Source facts
Action
The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.
Initiated
2019-04-19
Root cause
Nonconforming Material/Component
Terminated
2020-05-29
Distribution
The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.
Recall number
Z-0821-2020
Recall status
Terminated
CDRH record ID
176923
FDA classified
2020-01-19
Recall event number
83990