Z-0822-05: LifeScan brand OneTouch FastTake Blood Glucose Meter

Recall record Z-0822-05 for Lifescan Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0822-05
Company in FDA record
Lifescan Inc
FDA classification date
May 17, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

User may inadvertently change the Units of Measure from mg/dL to mmol/L and the blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia.

Source facts

Action
On 4/11/05, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions on the correction. Owner's booklet: Firm contacted all accounts by letters of 11/21/03, 2/4/2004, 2/17/04 and 2/9-10/2004 providing important information regarding the owner's booklet.
Initiated
2005-04-11
Root cause
Other
Terminated
2007-01-24
Distribution
Distribution of the product is worldwide. ***Distribution in the United States was between January 15, 2001 for OneTouch Ultra Blood Glucose Meter; April 13, 1998 for OneTouch FastTake Blood Glucose Meter and November 30, 2001 for InDuo Blood Glucose Meter.*** The total number of quantities shipped to consignees is approx. 1.8 Million. Of these, Approx. 282,000 were shipped to Pharmacies and Healthcare Professionals (including Diabetes Educators, Primary Care Physicians, Endocrinologists), Approx. 800 were shipped to Mail Order companies that distribute OneTouch ¿ FastTake¿ and OneTouch ¿ Ultra¿ products, and Approx. 130,000 were shipped to E-mail End User Notifications to LifeScan database of OneTouch¿ Ultra¿, OneTouch¿ FastTake¿ and InDuo¿ product users. Ultra, InDuo, FastTake Meters Global Distribution include: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Caribbean, CEE, Chile, China, Columbia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latin America, Latin America, Lebanon, Malaysia, Malta, MEA, Mexico, Netherlands, Norway, Oman, Pasean, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia/Gulf Countries, Singapore, Slovakia, Slovenia, South Africa, Spain, Sub-Saharan Africa, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam, and Yemen. Owner's booklet: The recall product was distributed in the U.S. only. LifeScan manufactured and distributed OneTouch Ultra System Kits with the affected OneTouch Ultra Meter Owner's Booklets between March 31, 2003 and October 3, 2003.
Recall number
Z-0822-05
Recall status
Terminated
CDRH record ID
38282
FDA classified
2005-05-17
Recall event number
31640

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗