Open original FDA record ↗Recall record
- FDA identifier
- Z-0822-2012
- Company in FDA record
- Biomet Microfixation, Inc.
- FDA classification date
- Jan 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Biomet Microfixation, Inc., Jacksonville, FL recalled Part # 24-6546, 45mm Left Standard Mandible and Part # 24-6646, 45mm Left Standard Offset Mandible due to a lot of 24-6246 being etched & labeled as 24-6646 and the reverse, a lot of 24-6646 being etched and labeled as 24-6546.
Source facts
- Action
- Biomet Microfixation Inc sent an URGENT-Medical Device Recall notice to all consignees. The letter informs the customers to discontinue use of the items and return them to Biomet Microfixation.They also ask all customers to complete the inventory reconciliation sheet and return it by fax to the Global Complaint Coordinator. The letter also states that if the product was further distributed, the letter must be forwarded on to the firm's customers. For questions call 1-800-874-7711 or 904-741-4400, ext 468.
- Initiated
- 2010-04-07
- Root cause
- Employee error
- Terminated
- 2012-02-06
- Distribution
- Worldwide Distribution - USA (nationwide) in the state of Florida and in the countries Germany and Guatemala.
- Recall number
- Z-0822-2012
- Recall status
- Terminated
- CDRH record ID
- 106770
- FDA classified
- 2012-01-20
- Recall event number
- 60910