Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04 meters to healthcare professionals.
Source facts
Action
Bayer Healthcare sent "IMPORTANT RECALL INFORMATION" to all affected customers.The letter identified the product, problem, and actions to be taken by the customers. Contact Viterion Support at 1-800-866-0133 option 1 for questions relating to this issue.
Initiated
2012-11-15
Root cause
Software design
Terminated
2017-02-16
Distribution
Worldwide Distribution-USA (nationwide) and the United Kingdom.
Recall number
Z-0822-2013
Recall status
Terminated
CDRH record ID
115433
FDA classified
2013-02-14
Recall event number
64111